Product name Clean-Q
Model no. TBD1000
Product classification Clean-Q
Download Product specification 제품규격서  /  manual 사용설명서
Major characteristics 1. Real-time monitoring of contamination by microorganism, mold, and other organic matters
2. Existing method: 24~48 hours, Clean-Q: only 10 seconds
3. Korea’s firstever sanitation monitoring terminal made with indigenous technology(completed patent registration), Korean language support
4. Optimized system for checking and managing sanitation status 5. Lowest error rate in the industry
Product Inquiry

Gauging mechanism and application

Clean-Q is a portable ATP(Adenosine Tri-Phosphate) gauge. All living organisms(animal/plant, microorganism, yeast, mold, etc.) use ATP as source of energy for their life activities. ATP can act as an indicator for all living organisms in existence and its amount can be used to measure microorganisms including germs. Clean-Q is used to gauge ATP amount on surface and any indication of contamination by measuring microorganism on a relative basis requires cleaning and cleansing. Clean-Q should be used after cleaning to check if contaminants are removed. Potential sanitation issues or incidents can be counteracted by checking microorganism on surface.

Product specification

Category

Specification

Category

Specification

Specification 10 sec. Gauging range 1-9.999 RLU
Gauging limit 5 x 10-15mol ATP Sensing error rate ± 5%
Memory 2,500 data storage PC communications USB
Working humidity 20 - 80% Working temperature 5 - 40 ℃
Power Two AA alkaline Battery life Battery life